Ambient Conditions (all models)

<< Click to Display Table of Contents >>

Navigation:  Hardware Specification >

Ambient Conditions (all models)

Temperature

0 - 40 °C

Relative humidity
N10, N20, X30-xCP, X36-xCP

20 - 80% (non-condensing)

Relative humidity
N30, N35, X30-xFA, X36-xFA

0 - 100% (non-condensing)

Note: The above limits apply to operation and storage of the camera.

Calibration Stability

Temperature changes inevitably cause slight deformations of the camera housing and its internal components. The approximate measurement deviations under temperature changes is quantified as follows:
 

Relative Z-coordinate deviation
(varies with operating distance)
 

≤ 5e-5 / °C at 1400mm distance
≤ 2.5e-5 / °C at 300mm distance

Angular X deviation
 

≤ 2e-5 rad / °C

Angular Y deviation
 

≤ 5e-5 rad / °C

 
Example

Let's assume you are operating your camera at approximately 300mm distance. The deviations you should expect under an ambient temperature change of 10°C are approximately:
 

Z-deviation: ≤ 2.5e-5/°C * 300mm * 10°C = 0.075mm

X-deviation: ≤ 300mm * tan(2e-5/°C * 10°C) = 0.075mm

Y-deviation: ≤ 300mm * tan(5e-5/°C * 10°C) = 0.150mm